SC-P100 Stylus Contact Profilometer

Cartesian stylus profilometer. X axis guided by high precision linear guide, Z1 axis digital sensor, software evaluates the measured contour.

Request a Quote

The instrument uses a Cartesian coordinate method with a contact stylus. The X axis runs on a high-precision linear guide as its reference and the Z1 axis uses a digital sensor, so the coordinate points of the workpiece surface profile are mapped directly. The software then processes the raw sensor data mathematically and labels the required measurement items.

Measurement functions

  • Size: horizontal distance, vertical distance, linear distance, radius, diameter
  • Angle: horizontal angle, vertical angle, included angle
  • Position tolerance: parallelism, perpendicularity
  • Form tolerance: straightness, crown, arc profile
  • Construction: auxiliary points, auxiliary lines, auxiliary circles

Installation data

Total weight and volume are given on the specification table below.

Operating requirements

  • Power: AC 220 V plus or minus 10 percent, 50 Hz
  • Environment: temperature 10 to 30 degrees C, relative humidity below 85 percent, no obvious vibration source near the installation, dedicated earth line

Technical parameters

  • X axis travel L equals the guide travel distance
  • Z1 axis measuring height H
  • Straightness within plus or minus 1
  • Arc R equals the standard arc value
  • Levelling system: overall size 250 x 180 x 40 mm, adjustment range 15 mm
  • Rotary and angle setting system: 360 degree rotation, plus or minus 25 degree tilt
  • Vice clamping system: jaw width 50 mm, jaw depth 25 mm
Component Item Qty
Mechanical Section Main Instrument Frame 1
Granite Worktable 1
Linear Guide Drive 1
Motorised Column 1
Grating Sensor 1
Adjustment Table Levelling System Dimensions 250 x 180 x 40 mm, adjustment range 15 mm 1
Rotary and tilt system: 360 deg rotation, plus or minus 25 deg tilt 1
Vice clamping: jaw width 50 mm, jaw depth 25 mm 1
Calibration Standard 1
Stylus and Probe Dia. 3 x 21 mm, R25 um, 12 deg single-facet stylus 2
8 x 150 mm probe 1
Electrical Section Electrical Control System 1
Brand-name PC 1
HP inkjet printer 1
Measuring Software 1
Documents User Manual 1
Warranty Card 1
Packing List 1
Certificate 1
Acceptance Record 1
Tool Box 1
Item Parameter
Measuring Range X axis Drive 100mm
Z1 (sensor) 40mm
Column travel 500mm
Indication accuracy X axis (L = guide travel) ±(0.8+0.02L)μm
Z1 axis (H = measuring height) ±(0.8+0.05H)μm
Angle ≤±1 ‘
Arc (R = nominal arc value) ≤±(1.2+R/12)μm
X axis Straightness 0.8μm/100mm
Resolution 0.1μm
Sensor Type Grating Sensor
Resolution 0.05μm
Stylus 12 deg single-facet stylus, R 0.025 mm
Climbing angle Up 77 deg, down 88 deg
Measuring Mode Sensor movement
Motion Control Guide speed 0.02 mm to 6 mm/s
Column movement speed 0.1 mm to 10 mm/s
Measuring Speed 0.02 mm to 5.0 mm/s
Product Packing Total Weight (kg) Volume (CBM m³)
Lx Ly Lz Wt(kg) Lx Ly Lz Wt(kg)
920 720 1700 350 950 800 1750 50 400 0.88